How to handle troubleshooting and error messages of JWQ? Aug 08, 2022


Feature:

·Resolution up to 1/15000

·Single point and linear calibration available

·Hold function in peak, stability, or key operation

·Lower/upper limit weighing alarming function

·Zero range setting and filter setting for counting under various condition

·Accumulation, Accumulation Display and Accumulation clear functions

·RS-232/Relay/Blue Tooth/U-KEY connector/WP wireless print/U-disk module/ Wifi/Ethernet (optional)

·Ideal for applications in production, packaging, warehouse, inventory, shipping and receiving areas


Specifications:

Model

JWQ-3K

JWQ-6K

JWQ-15K

JWQ-30K

Capacity(kg)

3

6

15

30

Division(g)

0.2

0.5

1

2

Display screen

128X51mm

Dimension(mm)

275X290X110

Weighing units

kg, g, t, lb, oz, pcs, 斤(台斤,港斤,市斤)

Weighing pan(mm)

275x 190(Stainless Steel, and 1.0mm in thickness)

Power supply

AC 100~240V / rechargeable battery (6 V/ 4 AH)



How important is it to quickly troubleshoot electronic weighing scales?

Electronic weighing scales are delicate instruments that consist of many components. Problems with each part or even each wire can cause inaccuracies, large errors, and non-display of values. It is important to quickly determine the cause of the electronic scale failure and to quickly investigate the cause and restore it to use.


How to handle troubleshooting and error messages of JWQ?


Error Message

Problem

shootings

ERR0

Exceed the zero range

The item should be within 2% of full load

ERR2

Exceed the initial zero point

1. Check whether there are other alien articles on the scale pan, remove those articles.

2. LOAD CELL failure, which requires to be changed or to contact our Service.

ERR3

Exceed the A/D resolution range

1. Check whether it is A/D failure, if yes, please replace AD.

2. LOAD CELL failure, replacement is required or contact our Service.

ERR4

EEPROM failure

Re-sold EEPROM or contact our Service.

ERR5

Overload condition

Remove weight that is greater than the scale capacity from the pan.

ERR6

Exceeds the display range

--------------------------------------------

ERR7

Accumulated number of weights exceeds the display range

Delete the exceeding weights

ERR8

Weight limit value is higher than the full load value

Reset the weight limit value.

ERR9

Exceed tare or pre-tare range

The tare value should be over zero and less than or equal to full load.

ERR10

Wrong calibration weights

Place the right weights( the calibration value ≤ full load)